The radiation dose, clinical and anatomical implications of erect lumbar spine radiography: A single centre pre-post implementation evaluation
View/ Open
bradley_and_snaith_2023 (505.4Kb)
Download
Publication date
2023-07Rights
© 2023 Elsevier. Reproduced in accordance with the publisher's self-archiving policy. This manuscript version is made available under the CC-BY-NC-ND 4.0 license (http://creativecommons.org/licenses/by-nc-nd/4.0/)Peer-Reviewed
YesOpen Access status
openAccess
Metadata
Show full item recordAbstract
Lumbar radiographs remain a common imaging examination despite strategies to reduce their use. Many authors have demonstrated benefits in changing from traditional supine and recumbent lateral projections to a prone and/or erect orientation. Despite evidence of clinical and radiation dose optimisation, widespread adoption of these strategies has stalled. This article describes the single-centre implementation and evaluation of erect PA and lateral projections. This was an observational study pre- and post-implementation of an erect imaging protocol. Patient BMI, image field size, source image and source object distances and DAP were collected together with assessment of radiographic spinal alignment and disc space demonstration. Effective dose was calculated with organ specific doses. 76 (53.5%) patients were imaged in the supine AP and recumbent lateral position, 66 (46.5%) had erect PA and lateral radiographs. Despite the larger BMI of the erect cohort and similar field sizes, effective dose was lower in the PA position by an average of 20% (pVersion
Accepted manuscriptCitation
Bradley C and Snaith B (2023) The radiation dose, clinical and anatomical implications of erect lumbar spine radiography: A single centre pre-post implementation evaluation. Radiography. 29(4): 690-696.Link to Version of Record
https://doi.org/10.1016/j.radi.2023.04.013Type
Articleae974a485f413a2113503eed53cd6c53
https://doi.org/10.1016/j.radi.2023.04.013