A novel particle isolation method for tissue samples was developed and tested using particle-doped peri-articular tissues from ovine cadavers. This enabled sensitivity of the isolation technique to be established by doping tissue samples of 0.25 g with very low particle volumes of 2.5 µm3 per sample. Image analysis was used to verify that the method caused no changes to particle size or morphologies.Version
Published versionCitation
Patel J, Lal S, Wilshaw SP et al (2018) Validation of a novel particle isolation procedure using particle doped tissue samples. Data in Brief. 18: 1802-1807.Link to Version of Record