Spatial Interpolation Enables Normative Data Comparison in Gaze-Contingent Microperimetry
View/ Open
Main article (1.661Mb)
Download
Publication date
2016-10Keyword
MicroperimetryFundus perimetry
Gaze-contingent
Normative database
Age-Related Macular Degeneration (AMD)
Spatial Interpolation
Rights
© 2016 The Authors. Published by ARVO. This work is licensed under a Creative Commons Attribution 4.0 International License.Peer-Reviewed
YesOpen Access status
openAccessAccepted for publication
2016-09-09
Metadata
Show full item recordAbstract
Purpose: To demonstrate methods that enable visual field sensitivities to be compared with normative data without restriction to a fixed test pattern. Methods: Healthy participants (n = 60, age 19–50) undertook microperimetry (MAIA-2) using 237 spatially dense locations up to 13° eccentricity. Surfaces were fit to the mean, variance, and 5th percentile sensitivities. Goodness-of-fit was assessed by refitting the surfaces 1000 times to the dataset and comparing estimated and measured sensitivities at 50 randomly excluded locations. A leave-one-out method was used to compare individual data with the 5th percentile surface. We also considered cases with unknown fovea location by adding error sampled from the distribution of relative fovea–optic disc positions to the test locations and comparing shifted data to the fixed surface. Results: Root mean square (RMS) difference between estimated and measured sensitivities were less than 0.5 dB and less than 1.0 dB for the mean and 5th percentile surfaces, respectively. Root mean square differences were greater for the variance surface, median 1.4 dB, range 0.8 to 2.7 dB. Across all participants 3.9% (interquartile range, 1.8–8.9%) of sensitivities fell beneath the 5th percentile surface, close to the expected 5%. Positional error added to the test grid altered the number of locations falling beneath the 5th percentile surface by less than 1.3% in 95% of participants. Conclusions: Spatial interpolation of normative data enables comparison of sensitivity measurements from varied visual field locations. Conventional indices and probability maps familiar from standard automated perimetry can be produced. These methods may enhance the clinical use of microperimetry, especially in cases of nonfoveal fixation.Version
Published versionCitation
Denniss J and Astle AT (2016) Spatial Interpolation Enables Normative Data Comparison in Gaze-Contingent Microperimetry. Investigative Ophthalmology and Visual Science. 57: 5449-5456.Link to Version of Record
https://doi.org/10.1167/iovs.16-20222Type
Articleae974a485f413a2113503eed53cd6c53
https://doi.org/10.1167/iovs.16-20222