Comprehensive assessment of patient image quality and radiation dose in latest generation cardiac x-ray equipment for percutaneous coronary interventions

Gislason-Lee, Amber J.
Keeble, C.
Egleston, D.
Bexon, J.
Kenyelics, S.M.
Davies, A.G.
Publication Date
2017
End of Embargo
Supervisor
Rights
(c) 2017 Society of Photo-Optical Instrumentation Engineers (SPIE). One print or electronic copy may be made for personal use only. Systematic reproduction and distribution, duplication of any material in this publication for a fee or for commercial purposes, and modification of the contents of the publication are prohibited.
Peer-Reviewed
Yes
Open Access status
openAccess
Accepted for publication
2017-04-14
Institution
Department
Awarded
Embargo end date
Additional title
Abstract
This study aimed to determine whether a reduction in radiation dose was found for percutaneous coronary interventional (PCI) patients using a cardiac interventional x-ray system with state-of-the-art image enhancement and x-ray optimization, compared to the current generation x-ray system, and to determine the corresponding impact on clinical image quality. Patient procedure dose area product (DAP) and fluoroscopy duration of 131 PCI patient cases from each x-ray system were compared using a Wilcoxon test on median values. Significant reductions in patient dose (p ≪ 0.001) were found for the new system with no significant change in fluoroscopy duration (p ¼ 0.2); procedure DAP reduced by 64%, fluoroscopy DAP by 51%, and “cine” acquisition DAP by 76%. The image quality of 15 patient angiograms from each x-ray system (30 total) was scored by 75 clinical professionals on a continuous scale for the ability to determine the presence and severity of stenotic lesions; image quality scores were analyzed using a two-sample t -test. Image quality was reduced by 9% (p ≪ 0.01) for the new x-ray system. This demonstrates a substantial reduction in patient dose, from acquisition more than fluoroscopy imaging, with slightly reduced image quality, for the new x-ray system compared to the current generation system.
Version
Published version
Citation
Gislason-Lee AJ, Keeble C, Egleston D et al (2017) Comprehensive assessment of patient image quality and radiation dose in latest generation cardiac x-ray equipment for percutaneous coronary interventions. Journal of Medical Imaging. 4(20): 025501.
Link to publisher’s version
Link to published version
Type
Article
Qualification name
Notes